qEV10 / 35 nm Gen 2 Column - 1 Pack

Size range 35nm-350nm - 10 ml Sample loading - Ideal for large volume cell culture supernatant

The qEV10 / 35 nm is now available in the Gen 2 range. Gen 2 qEV columns are built with Izon's new proprietary, agarose resin which delivers a more purified extracellular vesicle (EV)-containing eluate. The qEV10 / 35 nm Gen 2 column is designed for separating EVs from a 10 mL biological sample and comes with Luer Lock fittings, which allow for many different configurations.

The qEV10 / 35 nm Gen 2 has an optimal recovery range of 35 nm to 350 nm and is compatible with the Automatic Fraction Collector (AFC). AFCs are used to adjust the purified collection volume (PCV) to provide reproducible isolation for downstream analysis.

Izon products are designed and manufactured under a quality system certified to ISO 13485:2016.

qEV10 Gen 2 is also available in the 70 nm series.

Discounts are available on bulk orders of qEV Isolation Columns. Please contact us to find out more.

Product Code: IC10-35

Files

Technical Features

Specifications

Optimal Separation Size

<110nm

Buffer Volume *

23,2 mL

Sample load volume

10 mL*

Column Volume

69,3 mL

Optimal fraction size

5 mL

Flush Volume

140 mL

PCV*

20 mL

Elution peak after buffer volume*

15 ± 5 mL

Operational temperature 

18 to 24°C

Buffer

PBS

Upper, lower and loading frit size

20 µm

pH stability working range

3-13

pH stability cleaning-in-place (CIP)

2-14

Shelf life

12 months (if stored correctly)

*Loading higher sample volumes results in a lower level of purity in the later vesicle volumes, greater overlap between protein and EV elution peaks, and a higher protein peak within the PCV. Loading lower sample volumes results in a higher dilution factor of the sample. The optimal recommended sample volume for purity on the qEV10 Gen 2 is 10 mL.

**Values based off analysis of human plasma samples.