Accuracy is 0.01% of reading + 0.006% of full scale when used with MPT-141 probe.
Temperature range for instrument and probe 0 to 50° C
Because the probe characterisation information is stored in the probe assembly itself, any Group3 probe can be used with any Group3 DTM. However for full accuracy the DTM-151 should be used with a probe that has temperature compensation circuitry.
Two buttons on the front panel are used to select the operation of the teslameter. The following modes can be selected for the display - magnetic field, peak hold of magnetic field, AC component of field, probe temperature.
Digital filtering (time averaging) can be enabled to suppress short term fluctuations.
Several teslameters (up to 31) with serial communications can be formed into a local communications loop, all of them talked to by the one serial port on the control computer. The system can be operated in a trigger mode, where the timing of the measurements by several teslameters can be synchronised
Internal switches select serial data format, baud rate, device address, string terminators, filtering, gauss or tesla units, data format, service request action, EOI action, and reset system to defaults
Two analog outputs are available. These are a buffered version of the raw probe signal, and are not corrected for linearity or temperature.
All models are available without display or front buttons, for true "Black box" requirements.
Panel mount versions are available, and panels with correct cut-outs and mounting hardware.
Digital hall effect teslameter - gaussmeter
The DTM-151 Digital Teslameter offers extremely accurate, high resolution measurement of magnetic flux densities
Application Fields
Technical Features